Analysis of Natural Vitamin E Drugs
Mar 20, 2021
Method name: Determination of Vitamin E-Gas Chromatography
Scope of application: This method uses high performance liquid chromatography to determine the content of vitamin E.
This method is suitable for vitamin E.
Principle of the method: The test substance and internal standard are made into methanol solution, enter the gas chromatograph for chromatographic separation, and the absorption value of vitamin E (C31H52O3) and the internal standard n-tridodecane are detected at a wavelength of 254nm with an ultraviolet absorption detector , Calculate its content.
Reagents:
1. Isooctane
2.N-hexane
3. n-pentanol
equipment:
1. Instrument
1.1, high performance liquid chromatograph
1.2, chromatographic column
Use silicone (OV-17) as the stationary phase, and the coating concentration is 2%, or use HP-1 capillary column (100% dimethyl polysiloxane) as the analytical column; the theoretical plate number is calculated based on the vitamin E peak Not less than 500 (packed column) or 5000 (capillary column), the resolution of the vitamin E peak and the peak of the internal standard substance should meet the requirements.
1.3, UV absorption detector
2. Chromatographic conditions
2.1. Mobile phase: n-hexane + n-pentanol = 997 3
2.2. Detection wavelength: 294nm
2.3. Column temperature: 20℃
Sample preparation:
1. Weigh the test product
Accurately weigh 20 mg of this product and place it in a brown bottle with stopper.
2. Preparation of reference solution
Accurately weigh 25mg of vitamin E reference substance and vitamin E reference substance, put them in a 100mL brown measuring flask, add 80mL of isooctane, avoid heating, sonicate for 1 minute to completely dissolve, and add isooctane to the mark, shake, rinse Nitrogen tightly plugged, protected from light, and stored below 0°C.
3. Preparation of internal standard solution
Take an appropriate amount of n-tridodecane, add n-hexane to dissolve and dilute to make a solution containing 1.0 mg per 1 mL.
4. Preparation of test solution
Precisely add 10 mL of the internal standard solution to the above-mentioned test product, close it tightly, and shake to dissolve it to obtain the test product solution.
5. Determination of correction factor
Take an appropriate amount of n-tridodecane, add n-hexane to dissolve (dilute to make a solution containing 1.0 mg per 1 mL) and 20 mg of vitamin E reference substance, accurately weigh it, place it in a brown stopper, and add 10 mL of internal standard solution accurately. Stopper, shake to dissolve, take 1~3μL into the gas chromatograph, calculate the correction factor.
Note: "Precision weighing" means that the weighing should be accurate to one thousandth of the weight taken. "Precision measuring" means that the accuracy of measuring the volume should meet the accuracy requirements of the volume pipette in the national standard.
Steps:
Precisely absorb the above-mentioned test solution and reference solution respectively, and inject 1~3μL into the gas chromatograph. Use an ultraviolet absorption detector to determine the absorption value of vitamin E and internal standard n-tridodecane at a wavelength of 254nm, according to the internal standard method Calculated by the peak area, it is obtained.